YJ-201A Anaesthesia machine has Integrated Integrated breathing circuit with APLvalve, excellent tightness. Replacing drive bellow is NOT required when operating Adult/pediatric anesthesia.Applied with 1 high-precision vaporizers. (with optional to Selectatec). 7 inch screen with no dead space
Send Inquiry
Chat Now
Description
Technical Parameters

YJ-201A Anesthesia machines used in ICU

Medical equipment Portable ICU Anesthesia Machine

Feature:

•Integrated Integrated breathing circuit with APLvalve, excellent tightness.

•Replacing drive bellow is NOT required when operating Adult/pediatric anesthesia.

•Applied with 1 high-precision vaporizers. (with optional to Selectatec).

•7 inch screen with no dead space.

•Real-time detection for loops, Paw, BPM, VT and compliance.

•Comprehensive monitoring capabilities include ventilation parameters monitoring.preset parameters, alarm indication and breathing waveform monitoring.

•ACGO available to connect with T/Bain circuit for open operation.

•Yoke system with up to 3 back up cylinders (A type with 11L size).

Technique Specifications

Physical specifications

Dimensions:

52*73*123 cm

Weight:

60kg

Casters:

Front wheels with lock

Screen:

LCD

Display:

7 inch

Applications:

Patient range:

Audlt / Peadiatric

Language:

ENGLISH

Pipeline supply

Gas configuration

O2,N2O

Pipeline input range

280-600kPa

Working specifications

Power:

AC 220 v ± 10%, 50 Hz ± 2%

Current:

500mAax

Battery Power

Battery Type

Lead-acid 12V, 4Ah

Run-time

Typical 30 mins+, up to 2 hrs

Modes of ventilation

IPPV

Intermittent Positive Pressure Ventilation

SIMV

Synchronized Intermittent Mandatory Ventilation

MANUAL

Manual

SIPPV(A/C)

Synchronized Intermittent Positive Pressure Ventilation

STAND-BY

STAND-BY

Ventilator parameter

Working type

Electronically controlled, pneumatic driven

Tidal Volume range

40-1500ml

Pressure range

5-60 cmH2O

Rispriation Rate Range

1-99bpm (1-40bpm under SIMV)

I:E Range

4:1-1:8

Ptrigger Range

-20-0cmH2O

Minute volume range

>18L/Min

SIGH range

0-5/100

System Standard

Hypoxic guard system

N2O cut-off valve, O2 concentration >25%

Safety valve

<12.5kPa

O2 flush

25-75L/min

Volume of CO2 absorber

2L

Flow meter

4 tubes, Cascade 0.1-15L/min O2


0.1-15L/min N2O

Vaporizer

1 (with optional for Selectetac), maximum 2 mounts(Isoflurane/Enflurane/Sevoflurane/Halothane)

Monitoring

Types

VT. MV. BPM. Ppeak, Pmean, Inspiratory Platform, FiO2(Optional)

Waveforms

Waveforms of P-T, F-T

Alarm Limits

Audio/Visual

VT too high/too low,MV too high/too low,airway pressure, FiO2(Optional)

Oxygen failure,AC power failure,Battery low