| RESP | Technique Specification |
| • Method: Imped ance between RA-LL or RALA Measurement Range: Adult:10 ~ 120 BrPM · Neonatal /Pediatric: 7~150 BrPM Resolution: 1 BrPM • Accuracy: 2 BrPM | SPO2 · Measuring Range: 0~ 100 % • Resolution: 1% •Accuracy:70%~100%(±2%) ·Pulse Rate: 20-250 BPM • Resolution: 1 BPM ·Accuracy:±3 BPM |
| TEMP | Technique Specification |
| · Measure and Alarm Range: 0~50 C • Resolution: 0.1C ·Accuracy:0.1C | ECG Lead Mode: 5 Leads (I, II, III, AVR, AVL.AVF, V Gain:2.5mm/mV, 5.0mm/mV,10mm/mV 20mm/mV Heart Rate: 15-300 BPM (Adult); 15- 350 BPM (Neonatal) Resolution: 1 BPM Accuracy:±1% Sensitivity>200uV(Peak to peak) ST measurement range: -2.0~+2.0 mV Accuracy:-0.8mV~+0.8mV:±0.02mV or ±10%, which is greater Other Range: unspecified Sweep speed: 12.5 mm/s,25mm/s 50mm/s Bandwidth: Diagnostic: 0.05~130 Hz Monitor: 0.5~40 Hz Surgery: 1~20 Hz |
| Optinal Parameters: | |
| ·1.Etco2 · 2.Nellcor SpO2 。3.Sun Tech NIBP | |
| NIBP | |
| Method:Oscillometry Measure mode: Manual, Auto, STAT Unit mmHg, kPa • Measure and alarm range: • Adult Mode SYS 40~280 mmHg DIA 10~225 mmHg MEAN 20~240 mmHg • Pediatric Mode SYS 40~220 mmHg DIA 10~160 mmHg MEAN 20~170 mmHg · Neonatal Mode SYS 40~135 mmHg DIA 10~100 mmHg MEAN 20~ 110 mmHg · Resolution: 1mmHg ·Accuracy:±5mmHg | |
| Standard Parameters: | |
| · ECG, RESP, TEMP, NIBP, SPO2, PR | |
| Packing: Single package size: 34.7*20.7*27.7cm Single gross weight: 2.6 KG 6 unit per carton, package size: 50*28.7*61cm, total gross weight 16 KG |