YJ-FC113 Syringe Pump

YJ-FC113 Syringe Pump

YJ-FC113 Syringe Pump Features •3.5-inch LCD screen,Multiple languages Customization •Wide range of flow rate:0.01-1800ml/h •Automatic recognition of syringe specifications(5ml、 10ml、 20ml、 30ml、 50/60ml) •Dual-CPU design, more secure for system operation •Comprehensive audible and visual alarm...
Send Inquiry
Chat Now
Description
Technical Parameters
YJ-FC113 Syringe Pump

Syringe Pump 2

Features

•3.5-inch LCD screen,Multiple languages Customization

•Wide range of flow rate:0.01-1800ml/h

•Automatic recognition of syringe specifications(5ml, 10ml, 20ml, 30ml, 50/60ml)

•Dual-CPU design, more secure for system operation

•Comprehensive audible and visual alarm function, adjustable volume, visible from 360

•Dynamic pressure detection technology

•Drug library mode, customized version(optional), weight mode

•Change flow rate without stopping the current task

•Emergency power supply

•Nursing call port through RS232/WiFi

•Stackable and upgradable for workstations, better space efficiency with integrated power supply

 

Powerful alarm system with 360 visual light signs, easy for identification.

Manual & automatic bolus, can be programmed according to clinical use;

Automatic anti-bolus to decrease unexpected clinical risk.

Top accuracy in the world: +-2%, flow rate increments by 0.04ml/h.

Easy calibration of syringe and volume etc, For more friendly use.

Drug library contains more than 50 drug names, and can be updated & programmed.

Three infusion modes including Rate, Time, Body-weight.

DPS (Dynamic pressure system)

Hot Tags: yj-fc113 syringe pump, China, suppliers, manufacturers, factory, low price

Technical specification:

Model

YJ-FC113

Applicable syringes

5ml,10ml,20ml,30ml,50ml/60ml

Infusion Modes

Rate Mode, Time Mode, Body-Weight Mode

Drug Name Display

More than 50,

Can be updated according to customer requirements or clinical use

Accuracy

+-2%

Time Preset

0-99h and 59min

Flow Rate step by

Increments at 0.01ml/h

Flow rate

5ml: 0.01-150ml/h

10ml: 0.01-400ml/h

20ml: 0.01-600ml/h

30ml: 0.01-900ml/h

50ml: 0.01-1800ml/h

60ml: 0.01-1800ml/h

Safetly

Double CPU to make sure higher standard infusion

Volume limit

0.1-9999.99ml

Purge/Bolus Rate/Top flow rate

5ml: 150ml/h

10ml: 400ml/h

20ml: 600ml/h

30ml: 900ml/h

50ml: 1800ml/h

60ml: 1800ml/h

Audible&visible alarms

syringe dislocation,

occlusion ,

near end,

pause overtime,

low battery,

battery exausted,

VTBI completion,

syringe empty,

motor malfunction,

incorrect syringe specification,

Circult Malfunction,

internal communication malfunction,

master CPU malfunction,

monitoring CPU malfunction.

KVO

Rate≥10ml/h: KVO=3ml/h,

Rate<10ml/h: KVO=1ml/h

Occusion pressure

High: 106.7kpa+-26.7kpa,

Middle: 66.7Kpa+-13.3Kpa

Low: 40.7kpa+-13.3kpa

Pause overtime alarm

1min 50s-2min

Power & Battery

20VA

AC:100V-240V 50HZ/60HZ

DC:11.1V

Ni-MH battery support from 7 to 10hours

Calssification

Class II, type CF, IPX4

Operation Condition

Temperature: +-5C--+40degree C,

Relative Humidity: 20%-90%

Atmospheric pressure 70Kpa-106Kpa

Optional Function

Event log with at least 2000 history records;DC

Dimensions

140x260x110mm

Weight

2.5kg