Correct operation method of electrolyte analyzer

Apr 29, 2021 Leave a message

⒈The instrument is turned on and enters the system self-check to check whether the functions of the main components are normal, such as the main board of the instrument, the printer, the liquid path detection (completed by the liquid detector), the distribution valve and the valve detector, etc., which can intelligently identify and judge the fault, automatically prompt.


⒉Enter the activation electrode program, with the electrode activation timing function, accurately grasp the activation time, so as to improve the service life of the electrode and ensure the stability of the electrode. The time is counted down to 30 minutes, you can press the NO button to exit the activation electrode program directly.


⒊Enter the main menu, first perform system calibration, which can automatically select base point and slope calibration (you can also choose to exit the non-calibration, which is convenient for maintenance and maintenance, but it is not possible to directly measure serum samples)


⒋Select quality control analysis. After more than 5 quality control tests, a quality control report can be automatically generated and printed, and the average value, standard deviation, and coefficient of variation of the number of quality control performed can be calculated.


⒌Intelligent liquid detection program to ensure accurate sampling and measurement, automatic prompts for the measurement process, your convenient guide, 24 hours standby, automatic maintenance in standby mode, automatic positive and negative flushing function, short liquid path, unique positive Automatic backwashing calibration and flushing pipeline system to prevent cross-contamination and pipeline blockage.


⒍30 hole position automatic sampling system, with quality control M hole position and waste liquid hole position W position. 30 specimens can be tested at a time.


7. Automatic printing and manual printing are optional, saving printing paper. Report sheet: Comprehensive patient information report, which can set and print the normal range value of the human body.


⒏Using American wellhead compression-type power pump tube to increase the service life of the pump tube.


⒐Test samples: whole blood, serum, plasma, urine, body fluid, cerebrospinal fluid, urine, animal serum, etc.


⒑ Measurement method: ion selective electrode (ISE) direct method.


⒒Test items: K+, Na+, Cl-, Ca++, PH, CO2