1. After the end of each dialysis, if there is no macroscopic contamination, the outside of the dialysis machine should be initially sterilized and 500 mg/L of a chlorine-containing disinfectant or other effective disinfectant should be used for cleaning.
2. If blood contaminates the dialysis machine, wipe the blood immediately with a disposable cloth containing a chlorine disinfectant at a concentration of 1 500 mg/L, and wipe the outside of the machine with a 500 mg/L chlorine-containing disinfectant.
Dialysis Machine external disinfection purposes
Ensure that the exterior of the dialysis machine is free from contamination and blood-stained, and regulate the daily behavior of the relevant personnel for disinfection.
7 points to master
1. The machine wiping cloth can be used as a machine to disinfect one machine; one machine can be discarded to prevent secondary pollution.
2. Blood contamination External to the Dialysis Machine should be wiped with a disposable rag containing 1500mg/L chlorine-containing disinfectant to remove blood stains, and then disinfect the exterior of the machine with a 500mg/L concentration of chlorine-containing disinfectant.
3. Pay attention to the corners of the dialysis machine and wipe the blind area.
4. On the back of the dialysis machine, wipe once a week.
5. Infected area must be extension, partition, machine-specific.
6. For each wipe of a machine, disinfect it with a quick hand sanitizer and then wipe the next machine to prevent cross-infection.
7. The external dialysis machine must be sterilized every shift.