Check before use:
1. Make sure that the equipment is placed properly, and is safe and reliable;
2. Check whether there are foreign objects (such as liquid medicine residue, etc.) in the power socket;
3. Confirm whether the sound and visual alarm function is normal during the self-test, and check whether there is any damage at the same time.
4. Choose the appropriate infusion set and gear for the proposed treatment;
5. Check whether the battery power is sufficient. When a "battery undervoltage alarm" appears or the battery cannot be turned on normally due to insufficient power, it should be immediately connected to the network power supply for charging.
During use:
1. Select the part of the patient's blood vessel that is not susceptible to squeezing;
2. When replacing the device, disconnect the connection with the patient to prevent incorrect dose input;
3. Make sure that the infusion line is not kinked;
4. During the infusion process, The infusion bag is suspended above the pump obliquely to prevent the liquid from leaking directly into the pump. The liquid medicine dripping on the equipment should be wiped off in time to prevent the liquid medicine from flowing into the equipment and affecting the normal operation of the equipment