Common Faults And Solutions Of Biochemical Analyzer

Mar 02, 2021Leave a message

1.When the power is turned on, the fan does not rotate and does not display.

Cause of failure: fuse blown

Treatment method: replace with a new fuse (2A)

2.Poor stability.

Cause of failure: unstable light source voltage

Treatment method: check whether the input voltage meets the requirements of the machine

3.Poor repeatability.

The cause of the failure: air bubbles are generated when the sample is sucked

Treatment method: rinse with absolute ethanol for 2-3 times, then rinse with distilled water for 1-9 times

4.The printer does not print.

Cause of failure: poor contact of the plug or lack of paper

Solution: Replace the printing paper

5.The operation buzzer does not sound, and the operation fails and freezes.

Cause of failure: due to external grid interference or misoperation

Solution: Shut down and restart to run

6.The boot display is cluttered.

Cause of failure: This phenomenon is generally caused by grid interference

Solution: Turn off the power switch, wait 10 minutes and then restart, install a better quality regulator

7.Pump rotation, no sample injection

The cause of the failure: the suction tube is blocked or the silicone tube is aging

Treatment method: replace the suction tube or silicone tube