Bidding parameters of high-end anesthesia machine

Jun 20, 2021 Leave a message

Bidding parameters of high-end anesthesia machine

1. Part of the requirements for anesthesia machine:


1. Closed, semi-closed, semi-open, fully open mode


2. Air supply source: O2, N2O, AIR 0.28Mpa~0.60Mpa


★★3. Electronic flowmeter: O2, N2O, AIR 0~10L/min double column display, interface setting,


4. Rapid oxygen supply: 25L/min--75 L/min


5. Low gas source alarm: Pneumatic sound alarm when O2 gas source pressure is too low


6. Independent auxiliary oxygen inhalation device, adjustable flow range 0-15 l/min, operation knob and flow tube integrated with the anesthesia machine are built into the host


7. One-time calibration at the factory, no need to calibrate again, reducing the cost of use.


2. Anesthesia ventilator part:


1. Scope of application: adults, children


2. Drive mode: pneumatic electric control mode


3. Bellows: rising bellows


4. The integrated breathing circuit can be disassembled and installed without any tools, and can withstand high temperature and high pressure disinfection at 134C.


Avoid cross-infection in the hospital.


    ★★5. Equipped with general modes for adults, children, and children, as well as a separate mode for children


    ★★6. Display: ≥12.1 inch color LCD display mode, touch screen


      ★7. Breathing mode: VCV, PCV, SIMV-VC, SIMV-PC, PSV, MANUL, STANDBY, SIGH, DEMO


★★8. Tidal volume: Setting range: 10~1600 mL, with dynamic compliance compensation to ensure accurate tidal volume delivery without being affected by fresh gas flow


9. Equipped with fresh gas isolation valve technology to ensure that the tidal volume transmission is not affected by the fresh gas flow


★10. Breathing rate: Setting range: 1~100 bpm


 ★11. Inhalation to exhalation ratio: Setting range: 10:1~1:10


 12. Percentage of inhalation pause: Setting range: 0~50%;


       13. Positive end-expiratory pressure (PEEP): Setting range: OFF, 3~30 cmH2O; adjustable within the range, supporting heart and newborn


Child surgery to prevent hemorrhage.


       14. Pressure support: Setting range: 5~30 cmH2O


     ★15. Inhalation pressure limit: Setting range: 1~70 cmH2O


       16. Pressure trigger: Setting range: Range: -20~20 cmH2O;


     ★17. Monitoring parameters: tidal volume, respiratory rate, positive end expiratory pressure (PEEP), minute ventilation, oxygen concentration, airway pressure (peak


           Pressure, platform pressure)


       18. Respiratory waveform: volume-time waveform, pressure-time waveform, flow rate-time waveform, pulmonary function loop (flow-volume


           Ring; capacity-pressure ring)


19.Liquid crystal display: can display the graphs, data, trends, etc. of the above parameters


       20. Alarm system: Tidal volume upper and lower limit alarm, respiratory rate upper and lower limit alarm, airway pressure alarm, oxygen concentration alarm,


           Ventilation alarm, low battery, power failure alarm, low oxygen alarm


       21. Built-in battery: ≥4 hours, when the power or oxygen supply is interrupted, the anesthesia machine will automatically


Switch to backup power or air source and continue to work without causing death of the patient due to equipment failure.


       22. When the oxygen supply pressure is lower than a certain value, it will display and give a hypoxic alarm


       23. When the oxygen supply pressure is lower than a certain value, the machine will automatically cut off the sub-nitrogen oxide flow


       24. There will be an audible alarm immediately after power failure, and a low priority alarm will be activated. The alarm time is >120s


       25. Airway pressure upper limit alarm adjustment range: 0.5 ~ 8.0kPa


       26. Lower limit of airway pressure alarm: 0 ~ 4.0Pa


3. Technical parameters of anesthetic vaporizer:


1. Two anesthetic evaporation tanks


2. The evaporator has good air tightness


3. Equipped with temperature, flow and pressure compensation to ensure accurate output concentration


4. The output concentration of anesthetics can be adjusted continuously


       5. Enflurane, isoflurane, sevoflurane are available


4. Safety device:


       With sound and light alarm, power failure alarm, insufficient oxygen, automatic cut off laughing gas


5. Basic configuration:


      1. One main frame of anesthesia machine


      2. An anesthesia ventilator


      3. Two anesthetic vaporizers (without dosing device, enflurane and isoflurane)


      4. One carbon dioxide absorption loop


      5. A set of accessories