Cardiac monitoring common alarm processing process

Oct 17, 2022 Leave a message

Cardiac monitoring common alarm processing process

info-353-329

I. Abnormal heart rate alarm (during Cardiac monitoring)

1,First of all, check whether the electrode placement is correct, whether the electrode is in good contact with the skin, whether the electrode has fallen off, etc., to exclude interference factors affecting Cardiac monitoring results.

2,Check whether the alarm value is set in accordance with the condition to avoid missing and invalid alarm in Cardiac monitoring.

3,Monitor the heart rate with stethoscope to verify and confirm abnormal heart rate detected by Cardiac monitoring.

4,Immediately notify the physician on duty and make preparations for resuscitation based on Cardiac monitoring prompts.

5,Timely and accurate implementation of medical advice, and actively cooperate with physicians for resuscitation treatment guided by Cardiac monitoring data.

6,Closely observe the reaction after medication and monitor the change of the patient's heart rate through Cardiac monitoring.

Second, when the abnormal heart rhythm alarm (in Cardiac monitoring)

1,Nurses master the ECG changes of common arrhythmias, especially malignant arrhythmias, to accurately judge Cardiac monitoring prompts.

2,First check whether the electrode is correctly placed, whether the electrode is in good contact with the skin, whether the electrode is dislodged, etc., and exclude interfering factors that may affect Cardiac monitoring results.

3,Immediately notify the physician on duty and make preparations for resuscitation according to Cardiac monitoring alarm information.

4,Timely and accurately execute medical prescriptions, and actively cooperate with physicians to carry out rescue treatment based on Cardiac monitoring data.

5,Closely observe the reaction after medication and monitor the change of the patient's heart rhythm via Cardiac monitoring.

Third, when the blood pressure is abnormal alarm

1,First of all, check whether the cuff position is appropriate, whether the tightness is suitable, whether there is air leakage, etc.

2,Check whether the setting of the alarm value is in line with the condition to avoid missed and invalid alarms.

3,Repeat the measurement to confirm abnormal blood pressure

4,Immediately notify the physician on duty and make preparations for resuscitation

5,Timely and accurate implementation of medical advice, and actively cooperate with the physician for resuscitation treatment

6,Closely observe the reaction after medication and monitor the change of the patient's blood pressure.

Four, oxygen saturation abnormal alarm

1,First of all, check whether the finger sleeve is correctly placed and whether there are interfering factors such as falling off.

2,Monitor the patient's circulation and respiration to confirm the presence of abnormal oxygen saturation

3,Adjust oxygen flow, remove respiratory secretions and keep the airway open

4,If not relieved, immediately notify the physician, make preparations for resuscitation and correct hypoxemia

5,Close observation of changes in the condition, monitor the improvement of blood saturation

NIBP underinflation

Failure phenomenon: blood pressure measurement times "cuff too loose" or cuff leakage, inflation pressure is always not filled up (below 150mmHg), can not be measured.

Check method:

(1) There may be a real air leak, such as the cuff, air guide tube and various joints, through the "air leak detection" can be determined.

(2) The patient mode selection is not correct, if you use the adult cuff but the monitor patient type is used for newborns, there may be this alarm.

Exclusion method: Replace the blood pressure cuff with a good quality or choose the right type of cuff.

NIBP measurement value is not accurate

Fault phenomenon: the measured blood pressure value deviation is too large.

Checking method: Check whether the blood pressure cuff is leaking, whether the connection with the blood pressure tube interface is leaking, or whether it is brought about by a difference under subjective judgment with the auscultation method?

Exclusion method: Use the NIBP calibration function. This is the only available criterion to check if the calibration value of the NIBP module is correct at the user's site, and the standard deviation of the factory-checked pressure of the NIBP is within 8 mmHg. If this is exceeded, the blood pressure module needs to be replaced.

Transcutaneous Oxygen Saturation Monitoring

The infrared probe of the transcutaneous oxygen saturation monitor is fixed to the patient's fingertip, and the percentage of oxygenated hemoglobin to hemoglobin is monitored to the pulsation of the small artery at the patient's fingertip.

2,Cautions:

①The probe should be fixed when used and the patient should be kept as quiet as possible to avoid alarm and non-display of results.

②The accuracy of the results can be affected when the peripheral circulation is poorly perfused such as severe hypotension and shock.

No blood oxygen value or intermittent blood oxygen value

Fault phenomenon: When measuring human oxygen saturation, the oxygen value is intermittent.

Check method:

①When doing long-term monitoring and surgery, whether the patient's arm is moving around, causing the blood oxygen value to be intermittent.

②Check whether the oxygen extension cord is bad.

Solution: Keep the patient as stable as possible. Once the oxygen value is lost due to hand movement, it can be considered normal. If the oxygen extension cord is bad, replace it with a new one.