Introduction to the structure and function of conventional gas anesthesia machine
The conventional animal anesthesia machine generally consists of four parts: 1 oxygen and other auxiliary gases and flow meters; 2 anesthetic agent volatilization tank; 3 mixed gas output system; 4 waste gas recovery system.
Oxygen and other auxiliary gases and flow meters
The oxygen concentration used for animal anesthesia is not clearly defined, and ordinary medical oxygen is generally used directly.
In addition, some anesthesia machines use N2O and oxygen dual cylinders for air supply, which can reduce the amount of anesthetics, but currently domestic anesthesia machines with dual cylinders are rarely used.
The gas cylinder can be a general-purpose oxygen cylinder, and the gas cylinder needs to be equipped with a gas pressure reducing valve. For the use of anesthesia machine, it is required that the output gas pressure of the gas cylinder is not higher than 0.4MPA. There is a special pressure reducing valve abroad, the output gas pressure is not adjustable, and it is fixed at 50PSI (<0.4MPA).
The gas flow meter is a device that controls the amount of gas entering the anesthesia volatilization tank. Generally, the range of the flow meter is 0-4L/min for the anesthesia machine for larger animals, and the range of the general flow meter is for the anesthesia machine used in rats and mice. It is 0-1L/min.
Anesthetic volatilization tank
The anesthetic agent volatilization tank is the core component of the animal anesthesia machine. The liquid anesthetic agent is vaporized in the volatilization tank, driven by oxygen, and output at a certain concentration. The volatilization tank generally has an accurate scale, the scale range is from 0-5%, and the adjustment accuracy is 0.5%. Different anesthetics use different volatilization tanks and cannot be mixed. Regular manufacturers of anesthesia machines leave the factory, and the volatilization tank will bring a sheet of the factory inspection data of the volatilization tank when leaving the factory, clearly marking the actual value of each concentration scale.
During the use of the volatilization tank, the anesthetic is added from the front of the tank and the dosing hole in the lower part. Normally, a volatilization tank can add about 100ml of anesthetic. After the anesthetic is added, you can observe through the transparent window in the lower part of the volatilization tank. The amount of anesthetic agent, under normal conditions of use, the amount of anesthetic agent should be kept between the upper and lower scales.
In the process of using the anesthesia machine, it is divided into two stages, induction anesthesia and maintenance anesthesia. In the induction anesthesia, the scale of the volatilization tank is generally set at 3-4%, and the time for maintaining anesthesia is set at 1-2%.
Anesthesia volatilization tank
The anesthetic agent volatilization tank is the core component of the animal anesthesia machine. The liquid anesthetic agent is vaporized in the volatilization tank, driven by oxygen, and output at a certain concentration. The volatilization tank generally has an accurate scale, the scale range is from 0-5%, and the adjustment accuracy is 0.5%. Different anesthetics use different volatilization tanks and cannot be mixed. Regular manufacturers of anesthesia machines leave the factory, and the volatilization tank will bring a sheet of the factory inspection data of the volatilization tank when leaving the factory, clearly marking the actual value of each concentration scale.
During the use of the volatilization tank, the anesthetic is added from the front of the tank and the dosing hole in the lower part. Normally, a volatilization tank can add about 100ml of anesthetic. After the anesthetic is added, you can observe through the transparent window in the lower part of the volatilization tank. The amount of anesthetic agent, under normal conditions of use, the amount of anesthetic agent should be kept between the upper and lower scales.
In the process of using the anesthesia machine, it is divided into two stages, induction anesthesia and maintenance anesthesia. In the induction anesthesia, the scale of the volatilization tank is generally set at 3-4%, and the time for maintaining anesthesia is set at 1-2%.
Mixed gas output system
The output system of the mixed gas is the animal's anesthesia output pipeline, which can be divided into an open-circuit system and a closed-circuit system according to the different objects used.
The open circuit system is relatively simple, and the anesthetic gas can be delivered to the animal breathing mask through a simple pipeline. Kent's open circuit system has made a big improvement on this basis. Kent’s open-circuit system adopts a casing structure. The inner pipeline transmits anesthetic gas to the experimental animal, and the outer pipeline is connected to an activated carbon filter. The pressure of the animal’s exhalation is used to transmit the excess anesthetic gas to the activated carbon filter. It can greatly reduce the leakage of anesthetic gas caused by the open circuit system. Generally, the anesthesia machine for rats and mice uses an open-circuit system.
The structure of the closed-circuit system is more complicated than that of the open-circuit system. The principle diagram is as follows. The characteristic is that there is a circulation loop, and there is a CO2 absorber in the loop. The animal's CO2 is absorbed, and the anesthetic gas continues to circulate in the loop, and the animal can inhale the loop. Anesthetic gas in. The closed circuit system can save the amount of anesthetic and reduce the leakage of anesthetic, which is suitable for long-term surgery.
The anesthesia induction chamber and anesthesia mask are devices for delivering anesthetic agents to experimental animals during respiratory anesthesia. The picture is shown in the figure below. The induction chamber is generally a transparent square device with an anesthetic gas inlet and an air outlet, while the anesthesia mask is It is designed to meet the requirements of the animal's face. Generally, different animals have different designs of anesthesia masks.
The anesthetics used in animal anesthesia machines are all clinically used anesthetics, and the absorption of the product by the human body will also produce obvious anesthetic effects. Generally, in order to reduce the harm of anesthetics to the operators, it is necessary to select a corresponding exhaust gas recovery system. For Kent's small animal anesthesia machine, since it has built-in an activated carbon filter device, there is no need to purchase a related waste gas recovery system. The common exhaust gas recovery system is shown in the figure below. It is similar in principle to the setting of a vacuum cleaner. It has a built-in activated carbon adsorption device. The anesthetic gas in the air is pumped to the activated carbon inside the machine for adsorption through a high-horsepower exhauster.