What do the dry and wet types of biochemical analyzers mean?

Mar 03, 2022 Leave a message

1. The principle of wet chemical method is based on the Lambert-Beer law. The reaction carrier is an aqueous solution. The incident light is absorbed by the colored reaction product and then weakened. The concentration of the measured substance is reflected by the absorbance.


Dry chemistry, also known as solid-phase chemistry, uses solid-phase reagent technology with multilayer thin films. As long as the liquid sample is directly added to a reagent carrier that has been solidified in a special structure, that is, a dry chemical reagent, the water in the sample is used as a solvent to dissolve the reagent solidified on the carrier, and then mixed with the component to be tested in the sample. A chemical reaction is carried out to carry out analysis and determination, and finally the concentration or activity of the object to be measured is obtained.

POCT DRY ANALYZER

dry chemistry vs wet chemistry


dry chemistry

wet chemistry

Detection method

reflection photometry

Transmittance photometry

reaction vector

solid medium

Cuvette or flow cell

piping system

No liquid pipeline

with liquid pipeline

theoretical basis

Kubelka Munk

Lambert Beer

maintain

simpler

More

reagent

For solid phase, most of them do not need calibration, and the stability period is long; for liquid, calibration is required, and the stability period is short.

POCT ANALYZER 5

2. Advantages of dry chemistry


High accuracy


Sensitivity and accuracy are similar to typical discrete instruments.


Fast detection


Multilayer film technology, instant measurement after sample addition, short detection time


Wide range of applications


Flexible usage scenarios, suitable for laboratory departments, blood banks in clinical departments, primary medical and health institutions, physical examination centers, etc.


Environmentally friendly and pollution-free


No need to wash, clean and environmentally friendly waste only reagent cards, easy to collect and dispose of


Small and lightweight


The instrument is simple in structure, small in size and light in weight


easy to use


No liquid pipeline


Storage at room temperature, no need to rewarm, long-term stability without electrode maintenance, no need to replace the electrode membrane


Influencing factors of dry chemical analysis technology


1. Quality control substances


2. Storage and use of dry tablet reagents


3. Working environment and temperature