The Working Of High Frequency Electric Knife

Aug 19, 2016 Leave a message

The working principle of high frequency electric knife

There are two main operating modes for high-frequency electrosurgical units: unipolar and bipolar.


Unipolar mode:

In the monopolar mode, a complete circuit is used to cut and coagulate the tissue. The circuit consists of a high-frequency generator in a high-frequency electrosurgical unit, a patient plate, connecting wires and electrodes. In most applications, the current passes through the patient through effective wires and electrodes, and then returns to the generator of the high-frequency electrosurgical knife from the patient's plate and its wires.

The heating effect of the high-frequency electric knife that can destroy the diseased tissue is not caused by the heating electrode or the knife tip, like an electric cautery. It gathers high-frequency current with high current density and directly destroys the tissue at a point in contact with the tip of the effective electrode. When the temperature of the tissue or cell in contact with or adjacent to the effective electrode rises to the denaturation of the protein in the cell, coagulation occurs. This precise surgical effect is determined by the waveform, voltage, current, tissue type and electrode It depends on the shape and size.

In order to avoid heating the tissue when the current leaves the patient and returns to the high-frequency electrosurgical unit, the patient plate in the monopolar device must have a relatively large contact area with the patient to provide low impedance and low current density channels . Some high-frequency electrosurgicals used in doctors' offices have low current and low density, and can not use patient plates, but most general-purpose high-frequency electrosurgicals use larger currents, so patient plates are needed.

The output system isolated from the ground makes the current of the high-frequency electric knife no longer need the auxiliary channel between the patient and the ground, thereby reducing the risk of burning the body that may be in contact with the grounding object. However, if the ground-based system is adopted, the risk of burns is greater than that of the insulated output system.

Bipolar mode:

Bipolar coagulation is to provide high-frequency electrical energy to the body tissue through the two tips of the bipolar forceps, so that the blood vessels between the two ends of the bipolar forceps are dehydrated and coagulated to achieve the purpose of hemostasis. Its range of action is limited to the two ends of the forceps, and the damage to the body tissues and the range of influence are much smaller than that of the unipolar method. It is suitable for sealing small blood vessels (diameter <4mm) and fallopian tubes. Therefore, bipolar coagulation is mostly used in more delicate operations such as brain surgery, microsurgery, ENT, obstetrics and gynecology, and hand surgery. The safety of bipolar coagulation is gradually being recognized, and its scope of use is gradually expanding.

high frequency electric knife