1. Artificial disinfection methods for anesthesia masks, pipes, joints, humidifiers, and breathing bags
Completely dismantle the connections of the external circuit of the anesthesia machine and carefully inspect the pipe for defects such as sputum, blood stains and other contaminants.
Before the disinfection, it should be cleaned as required, and it must be thoroughly cleaned after soaking with enzyme solution.
Wash the pipes and accessories soaked in an effective disinfectant solution. When soaking, soak all in the disinfectant solution. The pipeline must not be bent. The cavity should be filled with disinfectant solution.
Conditional hospitals can also use other methods of low temperature sterilization after cleaning.
2. Anesthesia masks, pipes, fittings, humidifiers, respirator bags, etc.
Medical personnel should wear necessary protective equipment such as masks, hats, protective glasses, gloves, etc. before cleaning and disinfection.
The parts of the external circuit of the anesthesia machine are completely disassembled, and the parts are placed in the manner described in the manufacturer's instructions for the disinfector.
If there are blood stains, blemishes, and other contaminants on the external circuit, they can be pre-treated with enzymes and soaked in the cleaning and disinfection machine.
Follow the instructions of the manufacturer of the disinfector and disinfector to select an appropriate procedure for cleaning and disinfecting.
After cleaning, disinfection, and drying are completed automatically, they are placed in a clean bag and stored dry.